Difference between revisions of "Gossen Metrawatt Metrahit 16I"

From sigrok
Jump to navigation Jump to search
 
(7 intermediate revisions by the same user not shown)
Line 2: Line 2:
| image          = [[File:Gossen Metrawatt Metrahit 16I small.png|180px]]  
| image          = [[File:Gossen Metrawatt Metrahit 16I small.png|180px]]  
| name            = Gossen Metrawatt Metrahit 16I
| name            = Gossen Metrawatt Metrahit 16I
| status          = in progress
| status          = supported
| source_code_dir = <!-- serial-dmm -->
| source_code_dir = gmc-mh-1x-2x
| counts          = 3100 (31000)
| counts          = 3100 (31000)
| categories      = CAT III 1000 V, CAT IV 600 V
| categories      = CAT III 1000 V, CAT IV 600 V
Line 12: Line 12:
}}
}}


The '''Gossen Metrawatt Metrahit 16I''' is a 3100 counts handheld digital multimeter with IR (RS232) connectivity and insulation measurement (500/1000 V).
The '''[[Gossen Metrawatt]] Metrahit 16I''' is a 3100 counts handheld digital multimeter with IR (RS232) connectivity and insulation measurement (500/1000 V).


The resolution is one digit higher on the data interface.
The resolution is one digit higher on the data interface.
Line 18: Line 18:
== Hardware ==
== Hardware ==
* '''Battery''':  ''9 V IEC 6 LR 61''  
* '''Battery''':  ''9 V IEC 6 LR 61''  
* '''Fuse''': None!
* '''Fuse''': None! (not required, no internal amps ranges)


* TODO.
* TODO.
Line 44: Line 44:
| 30 mV
| 30 mV
| 10 μV
| 10 μV
| ±(0,5% + 3)
| ±(0.5% + 3)
| mq V, unit V, mqflags DC
| mq V, unit V, mqflags DC
| Not covered by autorange, use manual range
| Input impendance >10 GΩ; not covered by autorange, use manual range


|-
|-
| 300 mV
| 300 mV
| 100 μV
| 100 μV
| ±(0,5% + 3)
| ±(0.5% + 3)
| "
| "
| "
| Not covered by autorange, use manual range


|-
|-
| 3 V
| 3 V
| 1 mV
| 1 mV
| rowspan="3" | ±(0,25% + 1)
| rowspan="3" | ±(0.25% + 1)
| "
| "
|
|
Line 77: Line 77:
| 600 V
| 600 V
| 1 V
| 1 V
| ±(0,35% + 1)
| ±(0.35% + 1)
| "
| "
|
|
Line 137: Line 137:
| 30 A
| 30 A
| 10 mA
| 10 mA
| rowspan="2" | ±(2,5% + 3) (>10 Digit)
| rowspan="2" | ±(2.5% + 3) (>10 Digit)
| mq Curr., unit A, mqflags AC,DC,RMS
| mq Curr., unit A, mqflags AC,DC,RMS
| rowspan="2" | Current clamp with 100 mV/A required, e.g. GMC WZ12B
| rowspan="2" | Current clamp with 100 mV/A required, e.g. GMC WZ12B
Line 150: Line 150:
| 30.00 Ω
| 30.00 Ω
| 10 mΩ
| 10 mΩ
| rowspan="2" | ±(0,5% + 3)
| rowspan="2" | ±(0.5% + 3)
| mq Res., unit Ω, mqflags -  
| mq Res., unit Ω, mqflags -  
|
|
Line 163: Line 163:
| 3.000 kΩ
| 3.000 kΩ
| 1 Ω
| 1 Ω
| rowspan="3" | ±(0,4% + 1)
| rowspan="3" | ±(0.4% + 1)
| "
| "
|
|
Line 182: Line 182:
| 3.000 MΩ
| 3.000 MΩ
| 1 kΩ
| 1 kΩ
| ±(0,6% + 1)
| ±(0.6% + 1)
| "
| "
|
|
Line 189: Line 189:
| 30.00 MΩ
| 30.00 MΩ
| 10 kΩ
| 10 kΩ
| ±(2,0% + 1)
| ±(2.0% + 1)
| "
| "
|
|
Line 195: Line 195:
|-
|-
| Ω Iso 500V
| Ω Iso 500V
| 1,600 GΩ
| 1.600 GΩ
| 1 kΩ - 1 MΩ
| 1 kΩ - 1 MΩ
| rowspan="2" | ±(3% + 2)
| rowspan="2" | ±(3% + 2)
Line 203: Line 203:
|-
|-
| Ω Iso 1000V
| Ω Iso 1000V
| 3,100 GΩ
| 3.100 GΩ
| 1 kΩ - 1 MΩ
| 1 kΩ - 1 MΩ
| mq Res., unit Ω, mqflags -  
| mq Res., unit Ω, mqflags -  
Line 210: Line 210:
|-
|-
| Diode/Cont.
| Diode/Cont.
| 3.0000 V–
| 3.000 V–
| 1 mV
| 1 mV
| ±(0,25% + 1)
| ±(0.25% + 1)
| mq V, unit V, mqflags DC,DIODE
| mq V, unit V, mqflags DC,DIODE
| Diode and cont. are identical on this device, just without beeper for diode.
| Diode and cont. are identical on this device, just without beeper for diode.
Line 220: Line 220:
| 30.00 nF
| 30.00 nF
| 10 pF
| 10 pF
| rowspan="3" | ±(1,0% + 3)
| rowspan="3" | ±(1.0% + 3)
| mq Cap., unit F, mqflags -
| mq Cap., unit F, mqflags -
|
|
Line 239: Line 239:
| 30.00 µF
| 30.00 µF
| 10 nF
| 10 nF
| ±(3,0% + 3)
| ±(3.0% + 3)
| "
| "
|
|
Line 248: Line 248:
| 0.01 Hz
| 0.01 Hz
| rowspan="4" | ±(0.1% + 3)
| rowspan="4" | ±(0.1% + 3)
| mq Freq., unit Hz, mqflags DC
| mq Freq., unit Hz
|
| The protocol does not allow to differ AC and DC.


|-
|-
Line 255: Line 255:
| 0.1 Hz
| 0.1 Hz
| "  
| "  
|
|  


|-
|-
Line 274: Line 274:
| 0.01 Hz
| 0.01 Hz
| rowspan="4" | ±(0.1% + 3)
| rowspan="4" | ±(0.1% + 3)
| mq Freq., unit Hz, mqflags AC
| mq Freq., unit Hz
|
| The protocol does not allow to differ AC and DC.


|-
|-
Line 360: Line 360:
[[Category:Device]]
[[Category:Device]]
[[Category:Multimeter]]
[[Category:Multimeter]]
[[Category:In progress]]
[[Category:Supported]]

Latest revision as of 20:33, 21 September 2016

Gossen Metrawatt Metrahit 16I
Gossen Metrawatt Metrahit 16I small.png
Status supported
Source code gmc-mh-1x-2x
Counts 3100 (31000)
IEC 61010-1 CAT III 1000 V, CAT IV 600 V
Connectivity Infrared (RS232)
Measurements voltage, current, resistance, capacitance, temperature, diode, continuity
Features autorange, data hold, bargraph
Website [1]

The Gossen Metrawatt Metrahit 16I is a 3100 counts handheld digital multimeter with IR (RS232) connectivity and insulation measurement (500/1000 V).

The resolution is one digit higher on the data interface.

Hardware

  • Battery: 9 V IEC 6 LR 61
  • Fuse: None! (not required, no internal amps ranges)
  • TODO.

Photos

Measurement functions and ranges

Measurement function Range Resolution Deviation sigrok result Remarks
V= 30 mV 10 μV ±(0.5% + 3) mq V, unit V, mqflags DC Input impendance >10 GΩ; not covered by autorange, use manual range
300 mV 100 μV ±(0.5% + 3) " "
3 V 1 mV ±(0.25% + 1) "
30 V 10 mV "
300 V 100 mV "
600 V 1 V ±(0.35% + 1) "
V≈ 3 V 100 μV ±(1.0% + 3) (> 10 Digit) mq V, unit V, mqflags AC,RMS
30 V 100 μV "
300 V 1 mV "
1000 V 10 mV "
V≃ 3 V 100 μV ±(1.0% + 3) (> 10 Digit) mq V, unit V, mqflags AC,DC,RMS
30 V 100 μV "
300 V 1 mV "
1000 V 10 mV "
A≈ (Clamp only)  30 A 10 mA ±(2.5% + 3) (>10 Digit) mq Curr., unit A, mqflags AC,DC,RMS Current clamp with 100 mV/A required, e.g. GMC WZ12B
100 A 100 mA "
Ω 30.00 Ω 10 mΩ ±(0.5% + 3) mq Res., unit Ω, mqflags -
300.0 Ω 100 mΩ "
3.000 kΩ 1 Ω ±(0.4% + 1) "
30.00 kΩ 10 Ω "
300.0 kΩ 100 Ω "
3.000 MΩ 1 kΩ ±(0.6% + 1) "
30.00 MΩ 10 kΩ ±(2.0% + 1) "
Ω Iso 500V 1.600 GΩ 1 kΩ - 1 MΩ ±(3% + 2) mq Res., unit Ω, mqflags - Altogether 4 ranges with different resolution
Ω Iso 1000V 3.100 GΩ 1 kΩ - 1 MΩ mq Res., unit Ω, mqflags - Altogether 4 ranges with different resolution
Diode/Cont. 3.000 V– 1 mV ±(0.25% + 1) mq V, unit V, mqflags DC,DIODE Diode and cont. are identical on this device, just without beeper for diode.
Cap. F 30.00 nF 10 pF ±(1.0% + 3) mq Cap., unit F, mqflags -
300.0 nF 100 pF "
3.000 µF 1 nF "
30.00 µF 10 nF ±(3.0% + 3) "
Hz= 300.0 Hz 0.01 Hz ±(0.1% + 3) mq Freq., unit Hz The protocol does not allow to differ AC and DC.
3.000 kHz 0.1 Hz "
30.00 kHz 1 Hz "
100.00 kHz 10 Hz "
Hz≈ 300.0 Hz 0.01 Hz ±(0.1% + 3) mq Freq., unit Hz The protocol does not allow to differ AC and DC.
3.000 kHz 0.1 Hz "
30.00 kHz 1 Hz "
100.00 kHz 10 Hz "
°C Pt100 –200.0 - +200.0 °C 0.1 °C 2 Kelvin + 5 mq Temp., unit °C, mqflags -
Pt100 +200.0 - +800.0 °C ±(1.0% + 5) "
Pt1000 –100.0 - +200.0 °C 2 Kelvin + 5 "
Pt1000 +200.0 - +800.0 °C ±(1.0% + 5) "
Event count Seems not to be present on data output.
Data Hold mqflags += HOLD Disabled while IR interface is active on  Matthias' device.
MIN mqflags += MIN Influences only display, via IR flag is set, but actual current value is delivered.
MAX mqflags += MAX Influences only display, via IR flag is set, but actual current value is delivered.

The column "sigrok result" contains in short form what the driver generates for the resp. data type.