Difference between revisions of "Agilent U12xxx series"

From sigrok
Jump to navigation Jump to search
(start documenting CONF: commands)
(complete CONF: table for U123xx models)
Line 123: Line 123:
| <code>CONF:VOLT:AC 600</code>
| <code>CONF:VOLT:AC 600</code>
| Any voltage mode
| Any voltage mode
| Switch to Volts AC mode in 600V range. Valid ranges are: 600, 60, 6, 0.6
| Switch to Volts AC mode and 600V range. Valid ranges are: 600, 60, 6, 0.6


|-
|-
| <code>CONF:VOLT:AC 600</code>
| <code>CONF:VOLT:AC 600</code>
| Any voltage mode
| Any voltage mode
| Switch to Volts AC mode in 600V range. Valid ranges are: 600, 60, 6, 0.6
| Switch to Volts AC mode and 600V range. Valid ranges are: 600, 60, 6, 0.6


|-
|-
Line 139: Line 139:
| Any current mode
| Any current mode
| Switch to DC current measurement mode with autoranging
| Switch to DC current measurement mode with autoranging
|-
| <code>CONF:CURR:AC 60u</code>
| Any current mode
| Switch to AC current measurement mode and 60uA range. Valid ranges are 10, 6 for A mode, and 600u, 60u for uA mode.
|-
| <code>CONF:CURR:DC 60u</code>
| Any current mode
| Switch to DC current measurement mode and 60uA range. Valid ranges are 10, 6 for A mode, and 600u, 60u for uA mode.
|-
| <code>CONF:FREQ</code>
| VZlow, AC voltage or any current mode
| Switch to frequency measurement mode. No manual ranges are available.
|-
| <code>CONF:CONT</code>
| Resistance
| Switch to continuity mode.
|-
| <code>CONF:RES</code>
| Resistance
| Switch to resistance mode with autoranging.
|-
| <code>CONF:RES 6M</code>
| Resistance
| Switch to resistance mode and 6MΩ range. Valid ranges are 60M, 6M, 600k, 60k 6k, 600
|-
| <code>CONF:CAP</code>
| Capacitance
| Switch to capacitance mode with autoranging.
|-
| <code>CONF:CAP 1000n</code>
| Capacitance
| Switch to capacitance mode and 1000nF range. Valid ranges are: 10m, 1000u, 100u, 10u, 1000n





Revision as of 17:21, 2 September 2014

The Agilent U12xxx series are handheld digital multimeters, ranging from inexpensive all-purpose meters to industrial meters.

Model Counts Supported
U1231A 6000 Yes
U1232A 6000 Yes
U1233A 6000 Yes
U1241A / U1241B 10000 Yes
U1242A / U1242B 10000 Yes
U1251A / U1251B 50000 Yes
U1252A / U1252B 50000 Yes
U1253A / U1253B 50000 Yes
U1271A 30000 No
U1272A 30000 No
U1273A 30000 No
U1273AX 30000 No

Is your model not supported yet? Get in touch! We need people to help us test other models in this series, and getting a new model supported is generally quick & easy.

Connection

All models communicate via an optically isolated serial connection, at 9600 bps, 8 bits, no parity, and 1 stop bit (9600/8n1). Connecting to the devices can be done either via USB (U1173A cable) or Bluetooth (U1177A interface).

Protocol

All models in this series use generally the same protocol, with only minor differences between models -- and none within the same range.

*IDN?

Sending the command *IDN? causes the device to return a comma-separated four-field string, compatible with the identical GPIB command. The four fields are vendor name, model name, serial number and version:

Agilent Technologies,U1232A,MY52020136,V1.00

*RST

Sending the command *RST causes the device to reset. The response is an asterisk followed by the current dial position, for example *1 indicates the first position.

SYST:BATT?

The command SYST:BATT? returns a value indicating the device's battery level. The value is in a different format depending on the model series:

The U123xx models return a percentage value, like this:

36%

The U124xx models return a value like this:

+1.04200000E+02

SYST:BEEP

The command SYST:BEEP TONE make the multimeter emit a beep.

SYST:BLIT

The commands SYST:BLIT 1 and SYST:BLIT 0 turn the screen backlight on and off. There is no response. The backlight timer will still run and turn the backlight back off again if it is enabled.

FETC?

The FETC? command returns the current measured value, as a floating point ASCII string, as in the following examples:

+9.25000000E-03
+0.00000000E+00
-1.01140000E+00
-9.10200000E-01

The sign characters are always present, and the value is always in normalized scientific format. This applies even when in a mode that uses integer numbers, 11 would be transmitted as +1.10000000E+01. If no valid measurement is available, i.e. the display shows "OL" or "-OL", the transmitted value is either +9.90000000E+37 and -9.90000000E+37 respectively.

CONF:

The CONF: commands can be used to change modes and ranges on the multimeter. It is not to be confused with the CONF? query described below.

For a given rotary switch position, only those CONF: commands work that switch to modes that can be reached by pressing the range or shift buttons.

U123xx models

Command Rotary Switch Setting Meaning
CONF:VOLT:AC Any voltage mode Switch to Volts AC mode with autoranging
CONF:VOLT:DC Any voltage mode Switch to Volts DC mode with autoranging
CONF:VOLT:AC 600 Any voltage mode Switch to Volts AC mode and 600V range. Valid ranges are: 600, 60, 6, 0.6
CONF:VOLT:AC 600 Any voltage mode Switch to Volts AC mode and 600V range. Valid ranges are: 600, 60, 6, 0.6
CONF:CURR:AC Any current mode Switch to AC current measurement mode with autoranging
CONF:CURR:DC Any current mode Switch to DC current measurement mode with autoranging
CONF:CURR:AC 60u Any current mode Switch to AC current measurement mode and 60uA range. Valid ranges are 10, 6 for A mode, and 600u, 60u for uA mode.
CONF:CURR:DC 60u Any current mode Switch to DC current measurement mode and 60uA range. Valid ranges are 10, 6 for A mode, and 600u, 60u for uA mode.
CONF:FREQ VZlow, AC voltage or any current mode Switch to frequency measurement mode. No manual ranges are available.
CONF:CONT Resistance Switch to continuity mode.
CONF:RES Resistance Switch to resistance mode with autoranging.
CONF:RES 6M Resistance Switch to resistance mode and 6MΩ range. Valid ranges are 60M, 6M, 600k, 60k 6k, 600
CONF:CAP Capacitance Switch to capacitance mode with autoranging.
CONF:CAP 1000n Capacitance Switch to capacitance mode and 1000nF range. Valid ranges are: 10m, 1000u, 100u, 10u, 1000n


CONF?

The CONF? command queries the device's current measurement configuration.

U123xx models

This can be either one, two or three comma-separated fields. The number of fields depends on the first field's value, but the second field always denotes the measurement mode's range and resolution, and the third field is the current type (AC or DC). For example, this indicates AC voltage in the 600mV range:

V,0,AC
Mode Resolution Current Meaning
V y y Volts
0 Range 600mV, resolution 0.1mV
1 Range 6V, resolution 1mV
2 Range 60V, resolution 0.01V
3 Range 600V, resolution 0.1V
MV y y Millivolts
1 Range 600mV, resolution 0.1mV
A y y Ampere
0 Range 6A, resolution 1mA
1 Range 10A, resolution 0.01A
UA y y Microampere
0 Range 60μA, resolution 0.01μA
1 Range 600μA, resolution 0.1μA
FREQ y y AC frequency
0 Always shows 0, even when the range is changed manually on the device. This is possibly a firmware bug.
RES y n Resistance
0 Range 600Ω, resolution 0.1Ω
1 Range 6kΩ, resolution 0.001kΩ
2 Range 60kΩ, resolution 0.01kΩ
3 Range 600kΩ, resolution 0.1kΩ
4 Range 6MΩ, resolution 0.001MΩ
5 Range 60MΩ, resolution 0.01MΩ
CAP y n Capacitance
0 Range 1000nF, resolution 1nF
1 Range 10μF, resolution 0.01μF
2 Range 100μF, resolution 0.1μF
3 Range 1000μF, resolution 1μF
4 Range 10mF, resolution 0.01mF
DIOD n n Diode

U124xx and U125xx models

The output consists of the meter's mode, followed by a space, and then the meter's upper and lower ranges in that mode, separated by a comma. The mode may optionally include a colon followed by a code denoting a submode relevant to that major mode. For example, this is the AC voltage in the 1000mV range:

VOLT:AC +1.000000E+00,+1.000000E-04
Mode Meaning
VOLT Voltage, submode can be AC
CURR Current, submode can be AC
FREQ Frequency counter
DIOD Diode check
CONT Continuity
RES Resistance
CAP Capacitance
SCOU Switch counter
CPER Current loop, submode can be e.g. 0-20mA or 4-20mA.
T1 or T2 Temperature, submode can be K or J, denoting the thermocouple type. Instead of range, CEL or FAR is given, denoting the measurement is in Celcius or Fahrenheit.

STAT

The STAT? command is used to query the current state of the device. The response is a quoted 21-byte string, like this:

"000000000110L00000000"

U123xx models

Byte Value Meaning
1 Max/min/avg mode
0 Off
1 On
2 Relative measurement. The last value returned from FETC? before this mode was enabled denotes the stored reference value.
0 Off
1 On
3 Trig hold-log
0 Off
1 On
4 Auto hold-log
0 Off
1 On
5 LED flashlight
0 Off
1 On
6 LCD backlight
0 Off
1 On
7 Smoothing
0 Off
1 On
8 Temp/aux mode. When on and rotary switch is in position 5 (capacitance), CONF? returns MV,1,DC, but the value returned from FETC? instead denotes the temperature, in the unit (F or C) according to the meter's configuration.
0 Off
1 On
9 Unknown (always 0)
10 Beep frequency
0 4.2kHz
1 3.8kHz (default)
2 3.4kHz
3 3.2kHz
4 Off
11 Auto power-off (APO)
0 Off
1 On
12 Unknown (always 0)
13 Unknown (always L)
14-15 Unknown (always 0)
16 Rotary switch position
0 V/Zlow
1 Voltage AC
2 Voltage DC
3 Resistance
4 Diode
5 Capacitance
6 Current
7 Microcurrent
17 Continuity mode. When enabled, the value returned from FETC? is either NAN (no continuity) or any real value to denote continuity.
0 Off
1 On
18 Unknown (always 0)
19 Battery low indicator
0 Battery OK
1 Battery low
20-21 Unknown (always 0)

U124xx models

Byte Value Meaning
1 Max/min/avg mode
0 Off
1 On
2 Relative measurement. The last value returned from FETC? before this mode was enabled denotes the stored reference value.
0 Off
1 On
3 Unknown
4 Unknown
5 Unknown
6 Current loop mode
0 0-20mA mode
1 4-20mA mode
7 Unknown (always I)
8 Hold mode.
0 Off
1 On
9 Unknown
10 Beep frequency
0 Off (no beeps)
C 300Hz
F 600Hz
1 1.2kHz
2 2.4kHz
11 Auto power-off (APO)
0 Off
1 On
12 LCD backlight
0 Off
1 On
13 Unknown (always L)
14 Unknown
15 Unknown
16 Rotary switch position
0 Voltage
1 Diode
2 Resistance
3 Capacitance
4 uA
5 mA
6 A
7 Temperature
17 Unknown
18 Unknown (always 1)
19 Unknown
20 Switch counter mode
0 increase count on rising flank (initially low)
1 increase count on falling flank (initially high)
21 Auto range
0 Off
1 On

U125xx models

Byte Value Meaning
1 Max/min/avg mode
0 Off
1 On
2 Relative measurement. The last value returned from FETC? before this mode was enabled denotes the stored reference value.
0 Off
1 On
3 dB mode
0 Off
m ?
V ?
4 Unknown
5 Peak hold mode
0 Off
1 On
6 Current percent
0 0-20mA
1 4-20mA
7 Triggerhold I/B ?
8 Triggered hold mode.
0 Off
1 On
9 Zero temperature compensation?
10 Beep selection ?
11 Auto power-off (APO)
0 Off
1 On
12 LCD backlight
0 Off
1 On
13 Unknown (always L)
14 Unknown (always 0)
15 Unknown
16 Rotary position ?
17 Output status active/standby ?
18 Unknown
19 Battery low indicator
0 Battery OK
1 Battery low
20 Frequency counter prescaler
0 No prescaling
1 Divide by 100
21 Auto range
0 Manual range
1 Auto range