From: Frank Stettner Date: Sun, 12 Feb 2023 12:10:43 +0000 (+0100) Subject: hp-3478a: Fix range strings for current and resistance mode X-Git-Url: https://sigrok.org/gitweb/?p=libsigrok.git;a=commitdiff_plain;h=99307b251f726c453b6cec6600adcbcc1945320f hp-3478a: Fix range strings for current and resistance mode The range strings for the resistance mode were missing the unit (R). The range strings for both current modes (DC and AC) were wrongly using volts instead of amps. --- diff --git a/src/hardware/hp-3478a/api.c b/src/hardware/hp-3478a/api.c index 19d96ee1..c21546f2 100644 --- a/src/hardware/hp-3478a/api.c +++ b/src/hardware/hp-3478a/api.c @@ -71,21 +71,21 @@ static const struct { {SR_MQ_VOLTAGE, SR_MQFLAG_AC, 2, "300V"}, /* -99 is a dummy exponent for auto ranging. */ {SR_MQ_CURRENT, SR_MQFLAG_DC, -99, "Auto"}, - {SR_MQ_CURRENT, SR_MQFLAG_DC, -1, "300mV"}, - {SR_MQ_CURRENT, SR_MQFLAG_DC, 0, "3V"}, + {SR_MQ_CURRENT, SR_MQFLAG_DC, -1, "300mA"}, + {SR_MQ_CURRENT, SR_MQFLAG_DC, 0, "3A"}, /* -99 is a dummy exponent for auto ranging. */ {SR_MQ_CURRENT, SR_MQFLAG_AC, -99, "Auto"}, - {SR_MQ_CURRENT, SR_MQFLAG_AC, -1, "300mV"}, - {SR_MQ_CURRENT, SR_MQFLAG_AC, 0, "3V"}, + {SR_MQ_CURRENT, SR_MQFLAG_AC, -1, "300mA"}, + {SR_MQ_CURRENT, SR_MQFLAG_AC, 0, "3A"}, /* -99 is a dummy exponent for auto ranging. */ {SR_MQ_RESISTANCE, 0, -99, "Auto"}, - {SR_MQ_RESISTANCE, 0, 1, "30"}, - {SR_MQ_RESISTANCE, 0, 2, "300"}, - {SR_MQ_RESISTANCE, 0, 3, "3k"}, - {SR_MQ_RESISTANCE, 0, 4, "30k"}, - {SR_MQ_RESISTANCE, 0, 5, "300k"}, - {SR_MQ_RESISTANCE, 0, 6, "3M"}, - {SR_MQ_RESISTANCE, 0, 7, "30M"}, + {SR_MQ_RESISTANCE, 0, 1, "30R"}, + {SR_MQ_RESISTANCE, 0, 2, "300R"}, + {SR_MQ_RESISTANCE, 0, 3, "3kR"}, + {SR_MQ_RESISTANCE, 0, 4, "30kR"}, + {SR_MQ_RESISTANCE, 0, 5, "300kR"}, + {SR_MQ_RESISTANCE, 0, 6, "3MR"}, + {SR_MQ_RESISTANCE, 0, 7, "30MR"}, /* -99 is a dummy exponent for auto ranging. */ {SR_MQ_RESISTANCE, SR_MQFLAG_FOUR_WIRE, -99, "Auto"}, {SR_MQ_RESISTANCE, SR_MQFLAG_FOUR_WIRE, 1, "30R"},