1 -------------------------------------------------------------------------------
2 README.devices
3 -------------------------------------------------------------------------------
4
5 This README contains various notes for users of libsigrok (or frontends
6 that are based on libsigrok) about device- and/or driver-specific issues.
7
8
9 Firmware
10 --------
11
12 Some devices supported by libsigrok need a firmware to be uploaded every time
13 the device is connected to the PC (usually via USB), before it can be used.
14
15 The default locations where libsigrok expects the firmware files are:
16
17   $SIGROK_FIRMWARE_DIR (environment variable)
18   $HOME/.local/share/sigrok-firmware
19   $prefix/share/sigrok-firmware
20   /usr/local/share/sigrok-firmware
21   /usr/share/sigrok-firmware
22
23 ($prefix is usually /usr/local or /usr, depending on your ./configure options)
24
25 For further information see the section below and also:
26
27   http://sigrok.org/wiki/Firmware
28
29
30 Per-driver firmware requirements
31 --------------------------------
32
33 The following drivers/devices require a firmware upload upon connection:
34
35  - asix-sigma: The ASIX SIGMA and SIGMA2 require various firmware files,
36    depending on the settings used. These files are available from our
37    'sigrok-firmware' repository/project under a license which allows us
38    to redistribute them.
39
40  - fx2lafw: Logic analyzers based on the Cypress FX2(LP) chip need the
41    firmware files from the 'sigrok-firmware-fx2lafw' repository/project.
42    The firmware is written from scratch and licensed under the GNU GPLv2+.
43
44  - hantek-6xxx: Certain oscilloscopes based on the Cypress FX2(LP) chip, such
45    as the Hantek 6022BE/6022BL, SainSmart DDS120, and Rocktech BM102, need the
46    firmware files from the 'sigrok-firmware-fx2lafw' repository/project.
47    The firmware is written from scratch and licensed under the GNU GPLv2+.
48
49  - hantek-dso: The Hantek DSO-2090 (and other supported models of the same
50    series of Hantek PC oscilloscopes) need firmware files.
51    These can be extracted from the vendor's Windows drivers using a tool
52    from our 'sigrok-util' repository/project.
53
54  - lecroy-logicstudio: The LeCroy LogicStudio requires FPGA bitstream files.
55    These can be extracted from the vendor's Windows software using a tool
56    from our 'sigrok-util' repository/project.
57    Additionally, it requires a Cypress FX2 firmware. This can be extracted
58    from the vendor's Windows software using another tool. Details:
59
60      http://sigrok.org/wiki/LeCroy_LogicStudio#Firmware
61
62  - saleae-logic16: The Saleae Logic16 needs a firmware file for the
63    Cypress FX2 chip in the device, as well as two FPGA bitstream files.
64    These can be extracted from the vendor's Linux application using a tool
65    from our 'sigrok-util' repository/project.
66
67  - sysclk-lwla:
68
69     - The Sysclk LWLA1034 requires various bitstream files.
70       These files are available from our 'sigrok-firmware' repository/project
71       under a license which allows us to redistribute them.
72
73     - The Sysclk LWLA1016 requires various bitstream files.
74       These can be extracted from the vendor's Windows drivers using a tool
75       from our 'sigrok-util' repository/project.
76
77 The following drivers/devices do not need any firmware upload:
78
79  - agilent-dmm
80  - appa-55ii
81  - arachnid-labs-re-load-pro
82  - atten-pps3xxx
83  - baylibre-acme
84  - beaglelogic
85  - brymen-bm86x
86  - brymen-dmm
87  - cem-dt-885x
88  - center-3xx (including all subdrivers)
89  - chronovu-la
90  - colead-slm
91  - conrad-digi-35-cpu
92  - demo
93  - fluke-dmm
94  - ftdi-la
95  - gmc-mh-1x-2x (including all subdrivers)
96  - gwinstek-gds-800
97  - hameg-hmo
98  - hp-3457a
99  - hung-chang-dso-2100
100  - ikalogic-scanalogic2
101  - ikalogic-scanaplus
102  - kecheng-kc-330b
103  - kern-scale
104  - lascar-el-usb
105  - link-mso19
106  - manson-hcs-3xxx
107  - maynuo-m97
108  - mic-985xx (including all subdrivers)
109  - motech-lps-30x
110  - norma-dmm
111  - openbench-logic-sniffer
112  - pce-322a
113  - pipistrello-ols
114  - rigol-ds
115  - scpi-pps
116  - serial-dmm (including all subdrivers)
117  - serial-lcr (including all subdrivers)
118  - teleinfo
119  - testo
120  - tondaj-sl-814
121  - uni-t-dmm (including all subdrivers)
122  - uni-t-ut32x
123  - victor-dmm
124  - yokogawa-dlm
125  - zeroplus-logic-cube
126
127
128 Specifying serial ports
129 -----------------------
130
131 Many devices supported by libsigrok use serial port based cables (real RS232
132 or USB-to-serial ones) to connect to a PC.
133
134 For all these devices, you need to specify the serial port they are connected
135 to (e.g. using the 'conn' option in sigrok-cli). It is not possible to scan
136 for such devices without specifying a serial port.
137
138 Example:
139
140  $ sigrok-cli --driver <somedriver>:conn=/dev/ttyUSB0 ...
141
142 The following drivers/devices require a serial port specification:
143
144  - agilent-dmm
145  - appa-55ii
146  - atten-pps3xxx
147  - brymen-dmm
148  - cem-dt-885x
149  - center-3xx (including all subdrivers)
150  - colead-slm
151  - conrad-digi-35-cpu
152  - fluke-dmm
153  - gmc-mh-1x-2x (including all subdrivers)
154  - hameg-hmo
155  - link-mso19
156  - mic-985xx (including all subdrivers)
157  - norma-dmm
158  - openbench-logic-sniffer
159  - rigol-ds (for RS232; not required for USBTMC or TCP)
160  - serial-dmm (including all subdrivers)
161  - serial-lcr (including all subdrivers)
162  - teleinfo
163  - tondaj-sl-814
164
165 The following drivers/devices do not require a serial port specification:
166
167  - asix-sigma
168  - brymen-bm86x
169  - chronovu-la
170  - demo
171  - fx2lafw
172  - hantek-dso
173  - ikalogic-scanalogic2
174  - ikalogic-scanaplus
175  - kecheng-kc-330b
176  - lascar-el-usb
177  - pipistrello-ols
178  - rigol-ds (USBTMC or TCP)
179  - saleae-logic16
180  - sysclk-lwla
181  - uni-t-dmm (including all subdrivers)
182  - uni-t-ut32x
183  - victor-dmm
184  - yokogawa-dlm (USBTMC or TCP)
185  - zeroplus-logic-cube
186
187
188 Specifying serial port parameters
189 ---------------------------------
190
191 Every serial device's driver has default serial port parameters like baud
192 rate, number of data bits, stop bits and handshake status. If a device requires
193 different parameters, pass them as option "serialcomm" with the driver name.
194 See libsigrok docs for the function serial_set_paramstr() for complete specs.
195
196 Example:
197
198  $ sigrok-cli --driver <somedriver>:conn=<someconn>:serialcomm=9600/7n1/dtr=1
199
200
201 Permissions of serial port based devices
202 ----------------------------------------
203
204 When using devices supported by libsigrok that use serial port based cables
205 (real RS232 or USB-to-serial ones) to connect to a PC, you need to ensure
206 that the user running the libsigrok frontend has (read/write) permissions to
207 access the serial port device (e.g. /dev/ttyS0, /dev/ttyUSB0, and so on).
208
209 You can use 'chmod' to apply permissions as you see fit, and/or 'chown' to
210 change the owner of the serial port device to a certain user or group.
211
212 For USB-to-serial based devices, we recommended using our udev rules file
213 (see below for details).
214
215
216 Permissions for USB devices (udev rules file)
217 ---------------------------------------------
218
219 When using USB-based devices supported by libsigrok, the user running the
220 libsigrok frontend (e.g. sigrok-cli) has to have (read/write) permissions
221 for the respective USB device.
222
223 On Linux, this is accomplished using either 'chmod' (not recommended) or
224 using the udev rules file shipped with libsigrok (recommended).
225
226 The file is available in contrib/z60_libsigrok.rules. It contains entries
227 for all libsigrok-supported (USB-based) devices and changes their group
228 to 'plugdev' and the permissions to '664'.
229
230 When using a libsigrok package from your favorite Linux distribution, the
231 packager will have already taken care of properly installing the udev file
232 in the correct (distro-specific) place, and you don't have to do anything.
233 The packager might also have adapted 'plugdev' and '664' as needed.
234
235 If you're building from source, you need to copy the file to the place
236 where your distro expects such files. This is beyond the scope of this README,
237 but generally the location could be e.g. /etc/udev/rules.d, or maybe
238 /lib/udev/rules.d, or something else. Afterwards you might have to restart
239 udev, e.g. via '/etc/init.d/udev restart' or similar, and you'll have to
240 re-attach your device via USB.
241
242 Please consult the udev docs of your distro for details.
243
244
245 Cypress FX2 based devices
246 -------------------------
247
248 Devices using the Cypress FX2(LP) chip without any specific USB VID/PID will
249 be enumerated with VID/PID 04b4:8613 (the default for "unconfigured FX2").
250 These are usually "FX2 eval boards" (that can also be used as LAs, though).
251
252 On Linux, the 'usbtest' driver will usually grab such devices, and they will
253 thus not be usable by libsigrok (and frontends).
254
255 You can fix this by running 'rmmod usbtest' as root before using the device.
256
257
258 UNI-T DMM (and rebranded models) cables
259 ---------------------------------------
260
261 UNI-T multimeters (and rebranded devices, e.g. some Voltcraft models) can
262 ship with different PC connectivity cables:
263
264  - UT-D04 (USB/HID cable with Hoitek HE2325U chip, USB VID/PID 04fa:2490)
265  - UT-D04 (USB/HID cable with WCH CH9325 chip, USB VID/PID 1a86:e008)
266  - UT-D02 (RS232 cable)
267
268 The above cables are all physically compatible (same IR connector shape)
269 with all/most currently known UNI-T multimeters. For example, you can
270 use either of the UT-D04 USB/HID cables or the UT-D02 RS232 cable with
271 the UNI-T UT61D multimeter.
272
273 When using the UT-D02 RS232 cable with any of the supported UNI-T DMMs,
274 you have to use the respective driver with a '-ser' drivername suffix
275 (internally all of these models are handled by the 'serial-dmm' driver).
276
277 You also need to specify the serial port via the 'conn' option, e.g.
278 /dev/ttyUSB0 (attached via a USB-to-serial cable) or /dev/ttyS0 (actual
279 RS232 port) on Linux (see above).
280
281 Finally, the user running the frontend (e.g. sigrok-cli) also needs
282 permissions to access the respective serial port (see above).
283
284 Examples (sigrok-cli):
285
286  $ sigrok-cli --driver uni-t-ut61e-ser:conn=/dev/ttyUSB0 ...
287  $ sigrok-cli --driver voltcraft-vc820-ser:conn=/dev/ttyS0 ...
288
289 When using any of the UT-D04 USB/HID cables you have to use the respective
290 driver _without_ the '-ser' drivername suffix (internally all of these models
291 are handled by the 'uni-t-dmm' driver).
292
293 You also need to specify the USB vendor/device IDs of the cable.
294 Autodetection is not possible here, since various other products use the
295 USB VID/PID of those cables too, and there is no way to distinguish them.
296
297 Since the UT-D04 cables are USB based (but don't use a USB-to-serial chip)
298 there is no need to specify a serial port via 'conn', of course.
299 However, the user running the frontend does also need to have permissions
300 to access the respective USB device (see above).
301
302 Examples (sigrok-cli):
303
304  $ sigrok-cli --driver uni-t-ut61e:conn=1a86.e008 ...
305  $ sigrok-cli --driver voltcraft-vc820:conn=04fa.2490 ...
306
307
308 UNI-T UT-D04 cable issue on Linux
309 ---------------------------------
310
311 The UNI-T UT-D04 cable with Hoitek HE2325U (or WCH CH9325) chip seems to have
312 a very specific problem on Linux. Apparently it requires to be put into
313 suspend (and woken up again) before it is usable. This seems to be a
314 Linux-only issue, Windows is not affected by this since apparently the
315 Windows kernel does this for every USB device, always.
316
317 Thus, if you want to use any of the UNI-T DMMs with this specific cable,
318 you'll have to run the following script (as root) once, every time you attach
319 the cable via USB. The script was written by Ralf Burger.
320
321 See also: http://erste.de/UT61/index.html
322
323   #!/bin/bash
324   for dat in /sys/bus/usb/devices/*; do
325     if test -e $dat/manufacturer; then
326       grep "WCH.CN" $dat/manufacturer > /dev/null && echo auto > ${dat}/power/level && echo 0 > ${dat}/power/autosuspend
327     fi
328   done
329
330
331 Enabling multimeter / data logger measurement output
332 ----------------------------------------------------
333
334 Some multimeters or data loggers will not start outputting measurement data
335 unless a certain action has been performed by the user beforehand. This is
336 usually mentioned in the vendor manual of the respective device, but here's
337 a short list for convenience:
338
339  - BBC Goertz Metrawatt M2110: Briefly press the "Start/Reset" button on the
340    interface panel on top.
341  - Digitek DT4000ZC: Briefly press the "RS232" button.
342  - Gossen Metrawatt Metrahit 1x/2x devices, driver gmc-mh-1x-2x-rs232:
343    - Power on the device with the "DATA" button pressed.
344    - Metrahit 2x devices must be configured for the respective interface type.
345  - Gossen Metrawatt Metrahit 2x devices, driver gmc-mh-2x-bd232:
346    - 'BD232' interface:
347       The multimeter must be configured for the respective interface type.
348    - 'SI232-II' interface ("PC Mode"):
349       The multimeter must be configured for interface type 'BD232' (all),
350       'SI232 online' (28-29S) or 'SI232 store' (22-26x). The interface must
351       be configured to the same baud rate as the host (default 9600).
352       Multimeter and interface must be configured to the same address.
353  - Norma DM950: If the interface doesn't work (e.g. USB-RS232 converter), power
354    on the device with "FUNC" pressed (to power the interface from the DMM).
355  - PCE PCE-DM32: Briefly press the "RS232" button.
356  - RadioShack 22-812: Press and hold "SELECT" and "RANGE" together.
357  - TekPower TP4000ZC: Briefly press the "RS232" button.
358  - Tenma 72-7750: Briefly press the "RS232C" button.
359  - UNI-T UT60G: Briefly press the "RS232C" button.
360  - UNI-T UT61B/C/D: Press the "REL/RS232/USB" button for roughly 1 second.
361  - UNI-T UT71x: Press the "SEND/-/MAXMIN" button for roughly 1 second.
362    Briefly pressing the "EXIT" button leaves this mode again.
363  - UNI-T UT325: Briefly press the "SEND" button (as per manual). However, it
364    appears that in practice you don't have to press the button (at least on
365    some versions of the device), simply connect the device via USB.
366  - V&A VA18B/VA40B: Keep the "Hz/DUTY" key pressed while powering on the DMM.
367  - Victor 70C/86C: Press the "REL/RS232" button for roughly 1 second.
368  - Voltcraft VC-830: Press the "REL/PC" button for roughly 2 seconds.
369  - Voltcraft VC-870: Press the "REL/PC" button for roughly 1 second.
370
371
372 ChronoVu LA8/LA16 USB VID/PIDs
373 ------------------------------
374
375 The ChronoVu LA8/LA16 logic analyzer is available in two revisions. Previously,
376 the device shipped with a USB VID/PID of 0403:6001, which is the standard ID
377 for FTDI FT232 USB chips.
378
379 Since this made it hard to distinguish the LA8/LA16 from any other device
380 with this FTDI chip connected to the PC, the vendor later shipped the
381 device with a USB VID/PID of 0403:8867.
382
383 The 'chronovu-la' driver in libsigrok supports both VID/PID pairs and
384 automatically finds devices with either VID/PID pair.
385
386
387 OLS
388 ---
389
390 The Dangerous Prototypes Openbench Logic Sniffer (OLS) logic analyzer
391 driver in libsigrok assumes a somewhat recent firmware has been flashed onto
392 the OLS (it doesn't need a firmware upload every time it's attached via USB,
393 since the firmware is stored in the device permanently).
394
395 The most recent firmware version that is tested is 3.07.
396
397 If you use any older firmware and your OLS is not found or is not working
398 properly, please upgrade to at least this firmware version. Check the
399 Dangerous Prototypes wiki for firmware upgrade instructions:
400
401  http://dangerousprototypes.com/docs/Logic_Sniffer_upgrade_procedure
402
403 Also, you need to specify a serial port for the OLS in the frontends, e.g.
404 using the 'conn' option in sigrok-cli, and you also need to have the
405 permissions to access the serial port (see above).
406
407 Example: